The frequency of operational qualification of a CE instrument depends on factors such as the manufacturer s recommended OQ interval, the required instrument performance, and the instrument s use. Note that preventive maintenance procedures such as lamp replacement may require repeating the qualification of the particular CE function [3]. The following sections describe the tests to perform and their associated limits that are intended to qualify a CE instrument for its general purpose. Troubleshooting tips for instruments that do not perform to the intended specifications are presented in Section 12.4. [Pg.190]

Many detectors track the number of hours the lamp is ignited. Although the lamp life may vary, the detector s meter reading can be a helpful guide for troubleshooting. Lamps can sometimes operate for more than 2000 h. To distinguish a lamp problem from air bubbles, one should stop the mobile phase flow. A lamp problem will persist when the flow is stopped, whereas, if the problem is due... [Pg.1664]

Do the troubleshooting if the alarm lamp re-illuminates after its extinguishments following the 6-second-flashing (or it keeps... [Pg.396]

General troubleshooting guidelines are provided for ultraviolet (UV) units. A low radiation intensity indicator is provided with all UV units. The lamp intensity should be greater than 80%. [Pg.324]