Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Isotope Faraday cups

Unlike the photoplate, the Faraday detector (or Faraday cup) is still very much in use today. The main reasons for its lasting popularity are accuracy, reliability, and mgged construction. The simplest form of Faraday detector is a metal (conductive) cup that collects charged particles and is electrically connected to an instrument that measures the produced current (Fig. 2.21b). Faraday cups are not particularly sensitive and the signal produced must in most applications be significantly amplified. An important application for Faraday detectors is precise measurements of ratios of stable isotopes [278]. See, for example, Section 2.2.7 and Chapter 11 for examples of applications and methods in which Faraday detectors are utilized. [Pg.67]

Table 1. Sample configuration of dynamic Hg isotope analysis in three sequences. Faraday cups are labeled L for low masses, H for high masses and Ax for the axial collector. The axial cup is the reference so its efficiency is assumed to be unity. Table 1. Sample configuration of dynamic Hg isotope analysis in three sequences. Faraday cups are labeled L for low masses, H for high masses and Ax for the axial collector. The axial cup is the reference so its efficiency is assumed to be unity.
Modem isotope ratio mass spectrometers have at least three Faraday collectors, which are positioned along the focal plane of the mass spectrometer. Because the spacing between adjacent peaks changes with mass and because the scale is not linear, each set of isotopes often requires its own set of Faraday cups. [Pg.26]

Multi-collection mass spectrometers can analyze isotope ratios in a static mode to eliminate the errors from beam instability. However, the static multi-collection method depends on the extent to which the collectors (e.g., Faraday cups) are identical and to the extent to which the gain of each collector is stable. An alternative approach is to use the so-called dynamic multi-collector mode, to cancel out beam instability, detector bias, and performing a power-law mass fractionation correction. The following descriptions are modified from the Finnigan MAT 262 Operating manual (Finnigan, 1992). [Pg.217]

Thermal-ionization mass spectrometers (TIMS) combine a hot-filament source with a magnetic-sector mass spectrometer. The mass spectrometers are operated at low to moderate mass-resolving power. A large number of elements can be measured with thermal ionization mass spectrometry. Special care is taken to purify the samples using ion exchange columns. Samples are loaded onto the filaments along with an emitter, and a typical run may take several hours. Modem systems have multiple collectors so that several isotopes can be measured simultaneously. High-precision measurements are done with Faraday cup detectors, but low-abundance isotopes can be measured on electron multipliers. Modem machines are capable of precisions of 0.1 to 0.01 permit. [Pg.532]

A multiple ion collector device is required for the simultaneous determination of separated ion beams in precise and accurate isotope ratio measurements in order to study, for example, isotope fine variation in Nature or during tracer experiments using enriched stable isotope tracers. In thermal ionization mass spectrometers or in ICP-MS, mostly a system of several Faraday cups (up to 16) and/or ion counters (electron multipliers) is applied. In the photographs in Figures 4.7 and 4.8 examples of multiple ion collector systems are shown from the mass spectrometers MC-ICP-MS... [Pg.111]

Sr). Over the past 30 years, lead and strontium isotope ratios have been measured with thermal ionization mass spectrometry (TIMS). Elemental salts are deposited on a filament heated to produce ionized particles, which are then sent into a mass spectrometer where they are detected by multiple Faraday cups arrayed such that ions of several masses are collected simultaneously. TIMS is capable of high precision isotope discrimination, but the instruments tend to be large and expensive, and extensive sample preparation is required prior to sample introduction. Newer ICP-MS-based technologies like multi-collector ICP-MS (especially laser ablation) circumvent some of the sample preparation issues while exploiting the precision of simultaneous mass discrimination, but they are still limited by the number and configuration of ion collectors. [Pg.299]

TIMS analysis was performed on a fully automated VG Sector 54 mass spectrometer with eight adjustable faraday cups and a Daly ion-counting photomultiplier system. Analysis was performed in static mode. Each sample was analyzed 50 times to ensure acceptable precision. The TIMS analysis was standardized by use of the NIST SRM981 common lead standard. Multiple analyses of the SRM981 standard were used to determine a fractionation correction of 0.12% per amu and an overall error 0.06% per amu. Errors between runs of the same sample were below 0.01% per amu. This level of precision is comparable to the archaeometry database for lead isotopes (8). [Pg.318]

Another large instrument with emphasis on high transmission at high resolution for isotopic measurements in geological applications is the IMS 1270 [60]. This instrument is similar in design to the IMS-3f and other Cameca f-series instruments but is much larger. It operates in either microprobe or microscope mode. Four auxiliary detector assemblies, containing an electron multiplier or Faraday cup plus the central beam primary detector, provide simultaneous detection of five isotopes. [Pg.171]

Palacz, Z.A., Haines, C.H., and Turner, P.J. (1996) Performance of a new design of Faraday cup. Abstr. Todai Intemat. Symp. Cosmochemistry and Isotope Geochemistry, Tokyo, 194. [Pg.326]

The Faraday cup was widely used in the beginning of mass spectrometry but all the characteristics of this detector mean that it is now generally used in the measurement of highly precise ratios of specific ion species as in isotopic ratio mass spectrometry (IRMS) or in accelerator mass spectrometry (AMS). To obtain a highly accurate ratio in such relative abundance measurements, the intensities of the two stable beams of specific ions are measured simultaneously with two Faraday cups. [Pg.177]


See other pages where Isotope Faraday cups is mentioned: [Pg.36]    [Pg.37]    [Pg.49]    [Pg.403]    [Pg.471]    [Pg.471]    [Pg.85]    [Pg.63]    [Pg.66]    [Pg.168]    [Pg.173]    [Pg.199]    [Pg.201]    [Pg.84]    [Pg.141]    [Pg.142]    [Pg.200]    [Pg.175]    [Pg.170]    [Pg.531]    [Pg.532]    [Pg.11]    [Pg.105]    [Pg.112]    [Pg.132]    [Pg.135]    [Pg.137]    [Pg.166]    [Pg.167]    [Pg.168]    [Pg.249]    [Pg.404]    [Pg.142]    [Pg.11]    [Pg.295]    [Pg.295]    [Pg.418]    [Pg.419]    [Pg.341]    [Pg.11]   
See also in sourсe #XX -- [ Pg.280 , Pg.281 ]




SEARCH



Faraday

Faraday cup

© 2024 chempedia.info