How to operate a flame atomic absorption spectrometer

You should only operate an FAAS system iiiuli t iliiecl supervision. The instrument should be located under a fume extraction hood. The spectrometer requires approximately 20 min to warm up before switching the gases on and using the instrument. [Pg.177]

Adjust the operating wavelength and slit width of the monochromator. This is done by cnnsiilling standard operating conditions, e.g. for load see Table 27.2. [Pg.177]

Decide what wavelength is to be used for the analysis. For lead llie maximum sensitivity is achieved by selecting 217.0 nm. [Pg.177]

Adjust the gain control until the energy motor reading reaches a maximum. [Pg.177]

Adjust the wavelength selector for maximum signal reading. You aro now ready to ignite the air-acetylene flame. [Pg.177]


See other pages where How to operate a flame atomic absorption spectrometer is mentioned: [Pg.177]    [Pg.177]    [Pg.371]   