Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

High-mass instrumentation

Peak matching can be done on quadrupole and magnetic-sector mass spectrometers, but only the latter, particularly as double-focusing instruments, have sufficiently high resolution for the technique to be useful at high mass. [Pg.274]

SiiH2 are completely separated from the peak. Quadrupole instruments are not usually capable of such high mass resolution. [Pg.544]

The choice of mass spectrometer for a particular analysis depends on the namre of the sample and the desired results. For low detection limits, high mass resolution, or stigmatic imaging, a magnetic sector-based instrument should be used. The analysis of dielectric materials (in many cases) or a need for ultrahigh depth resolution requires the use of a quadrupole instrument. [Pg.548]

The four-sector mass spectrometer is the ultimate in MS-MS instrumentation and consists of two high-resolution mass spectrometers in series. The strength of these instruments is in terms of their high-mass and high-resolution capabilities for both precursor-ion selection and product-ion analysis. Their cost, however, precludes their primary use for LC-MS and therefore they will not be considered any further here [12]. [Pg.67]

The mass range requirement invariably means that FAB is used in conjunction with a magnetic sector instrument. Conventional detectors, such as the electron multiplier, are not efficient for the detection of large ions and the necessary sensitivity is often only obtained when devices such as the post-acceleration detector or array detector are used. Instruments capable of carrying out high-mass investigations on a routine basis are therefore costly and beyond the reach of many laboratories. [Pg.157]

For polymer/additive analyses neither a very high mass range nor an ultrahigh mass resolution is required isobaric additive ions are not frequent. It is therefore not surprising that tandem sector instruments have not found wide application for polymer/additive identification. [Pg.356]

In recent years several new instruments have been developed based on different mass-spectrometer principles. Two different categories of ICP-MS instruments are currently commercially available low-resolution instruments (using either QMS, ITMS or ToF-MS) and focusing high-resolution instruments (DFS, FTMS). Selected specifications for these two categories are shown in Table 8.63. Both the quadrupole-based and the double-focusing instruments allow a sequential multielement measurement, whereas ICP-ToFMS allows... [Pg.655]

Lasers have advanced the analytical use of mass spectrometers to characterise additives in polymers, and routine application of MALDI is no longer limited to high molecular masses only. MALDI can now clearly produce isotopically resolved mass spectra of small molecules (<800 Da) in an L-ToF instrument, which can be used successfully for the characterisation of molecules of different chemical classes. High mass resolution with an improvement of mass accuracy to... [Pg.702]

The ability to detect small genetic changes becomes more difficult as mass increases. There is further an upper mass range where analysis is impractical. For low-resolution instruments this limit is around a 100 mer. Thus the mass has to be minimized or a high-resolution instrument employed. Alternatively, the smaller the piece of DNA analyzed, the more it chemically resembles a primer or nucleotide monomer thus separation of the two during cleanup is difficult to do. If the primers and nucleotides are not removed, they can provide a massive background on MS analysis or inhibit ionization of the PCR product by preferential ionization. Thus for practical reasons it is extremely difficult to employ a PCR product below a 40 to 50mer for direct ESI MS or ESI MS-MS analysis. [Pg.31]

The second serious problem of a high transmission instrument relates to the presence of molecular fragments. These molecules, which originate in the source, have nearly the same mass as the wanted ions. For 14C measurements, the mass difference between this atom and the molecule 12CH2 is approximately 0.1 percent. The resolution needed to separate these components is about 6,000, because there may be 109 12CH2 molecules for each 14C atom. [Pg.54]

The m/z values of peptide ions are mathematically derived from the sine wave profile by the performance of a fast Fourier transform operation. Thus, the detection of ions by FTICR is distinct from results from other MS approaches because the peptide ions are detected by their oscillation near the detection plate rather than by collision with a detector. Consequently, masses are resolved only by cyclotron frequency and not in space (sector instruments) or time (TOF analyzers). The magnetic field strength measured in Tesla correlates with the performance properties of FTICR. The instruments are very powerful and provide exquisitely high mass accuracy, mass resolution, and sensitivity—desirable properties in the analysis of complex protein mixtures. FTICR instruments are especially compatible with ESI29 but may also be used with MALDI as an ionization source.30 FTICR requires sophisticated expertise. Nevertheless, this technique is increasingly employed successfully in proteomics studies. [Pg.383]

Other instruments include the Calvet microcalorimeters [113], some of which can also run in the scanning mode as a DSC. These are available commercially from SETARAM. The calorimeters exist in several configurations. Each consists of sample and reference vessels placed in an isothermally controlled and insulated block. The side walls are in intimate contact with heat-flow sensors. Typical volumes of sample/reference vessels are 0.1 to 100 cm3, The instruments can be operated from below ambient temperatures up to 300°C (some high temperature instruments can operate up to 1000°C). The sensitivity of these instruments is better than 1 pW, which translates to a detection limit of 1 x 10-3 W/kg with a sample mass of 1 g. [Pg.63]

TOF analyzer it is critical for the mass resolution that the secondary ions are ejected at a precisely defined time. This means that the primary ion pulse should be as narrow in time as possible, preferably < 1 ns. At the same time maximum lateral resolution is desired. Unfortunately, there is a trade-off between these two parameters if the primary ion intensity is not to be sacrificed [122], Therefore, TOF-SIMS instruments have two modes of operation, high mass resolution and high lateral resolution. An advantage with the pulsed source is that an electron flood gun can be allowed to operate when the primary ion gun is inoperative. Thus, charge-compensation is effectively applied when analyzing insulating materials. [Pg.33]


See other pages where High-mass instrumentation is mentioned: [Pg.523]    [Pg.552]    [Pg.563]    [Pg.612]    [Pg.612]    [Pg.622]    [Pg.411]    [Pg.13]    [Pg.50]    [Pg.58]    [Pg.62]    [Pg.100]    [Pg.104]    [Pg.32]    [Pg.36]    [Pg.36]    [Pg.38]    [Pg.72]    [Pg.40]    [Pg.44]    [Pg.381]    [Pg.401]    [Pg.505]    [Pg.511]    [Pg.542]    [Pg.656]    [Pg.47]    [Pg.288]    [Pg.57]    [Pg.46]    [Pg.48]    [Pg.70]    [Pg.97]    [Pg.244]    [Pg.338]    [Pg.372]   
See also in sourсe #XX -- [ Pg.36 ]




SEARCH



High instrumentation

High-mass

© 2024 chempedia.info