Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Gradient Calibration

The function /, describing the relationship between concentration and delay time t, is dependent on the dispersion process taking place in the FIA channel and, therefore, on channel geometry, volume, and flow velocity. In the simplest case, that is, absence of chemical reactions, and when the flow channel conforms with the model of one well-stirred tank (cf. Chapter 3), the concentration-time matrix of the falling edge of the curve is described by  [Pg.51]

If a mixing chamber is not used, as is the case in most FI A systems, it should be noted that the injection process itself results in the formation of a concentration gradient similar to the one stirred tank [183] and that the thus formed sample zone will be transformed gradually into a Gaussian shape as the number of mixing stages increase (cf. Fig. 2.66, and Chapter 3). Calibration by means of known standards is then the only practical way to establish the matrix of t = f(C ). [Pg.52]

3 Reaction Rate Measurement by Stopped Flow on the Gradient [Pg.52]

Except for FIA titrations, the stopped-flow method [20, 264] is, the most frequently used gradient technique. The most practical approach to reaction rate measurement is to halt the reaction mixture within the observation volume of the flowthrough detector by stopping the flow and to measure the change of the signal with time. This is readily accomplished by the FIA stopped-flow technique (Chapter 4.3), which relies on a combination of a suitable dispersion of the sample material within the reagent stream and on the subsequent reaction rate measurement during the stopped-flow period. [Pg.52]

The recorded signal contains a portion of the reaction-rate curve, the slope of which is the basis of the calibration curve for a fixed-time kinetic assay. A closer look at the dispersing sample zone (Fig. 2.21, left) shows [Pg.52]


The CHI index is reportedly a relevant parameter in quantitative structure-activity relationship (QSAR) studies [41]. With this approach, log P could be determined in the range -0.45more than 25000 compounds with excellent reproducibility (within 2 index units) and reported in a GlaxoSmithKline database [11]. Two main drawbacks were identified using this approach (i) the assumptions used in Ref [7], i.e. that S is constant for all compounds and that the system dwell volume is excluded in calculations, yield some discrepancies in the resulting log P, and (ii) the set of gradient calibration... [Pg.342]

Precision in the measured diffusivities is limited by the reproducibility of the echo height measurements and by field gradient calibration. Under favorable circumstances, both of these can be kept below about 1 %, although for less time-consuming and routine measurements, 2-3 % random uncertainty is more typical. If comparisons among different samples are more important than comparisons with other work on similar samples, then calibration error is quasisystematic and secondary. In that case, perhaps 4% calibration uncertainty may be acceptable, requiring measurement of G or G0 to within 2% (see Eqs. 1, 2). [Pg.8]

Essentially there are two approaches to gradient calibration, one based on the measurement of molecular diffusion, the other using an image of a suit-... [Pg.100]

The weight of all the atoms or ions in the unit cell can be calculated if the unit cell dimensions and the density of the crystal are known. This means that crystal density measurements are important in the early stages of a crystal structure analysis. A vertical column with a density gradient, calibrated with crystals of known density, is used to measure protein densities. These are determined by where along the column the crystal setdes and so that, by interpolation, its density can be estimated (39). This method must be done with care to ensure that the measurement is good. The density of a crystal is given by... [Pg.33]

Figure 6.17 shows the experimental set-up proposed by Tyson et al. [86] for gradient calibrations. It consists of two reservoirs holding water and a concentrated standard solution of the metal Ion, linked via a three-way valve which allows the two solutions to be sent separately to the nebulizer. [Pg.187]

Figure 2.20. The principle of gradient calibration, showing on the left-hand side a conventional calibration run where standards covering 25%, 50%, 75%, and 100% of the range were injected each in triplicate, and on the right-hand side a single injection of the 100% standard, which was recorded at a high chart speed. Via a conventional calibration line (r = 1000), delay times t, t2, ts, and t4 are then identified, corresponding to known levels of originally injected standards. Figure 2.20. The principle of gradient calibration, showing on the left-hand side a conventional calibration run where standards covering 25%, 50%, 75%, and 100% of the range were injected each in triplicate, and on the right-hand side a single injection of the 100% standard, which was recorded at a high chart speed. Via a conventional calibration line (r = 1000), delay times t, t2, ts, and t4 are then identified, corresponding to known levels of originally injected standards.
The reality can well be approximated by keeping an unknown systematic error constant in each simulation, but vary it between simulations. Examples are probe diameters, temperatures and temperature gradients, calibration-uncertainties of material measures, etc. [Pg.855]

Medium Calibration (gradient calibration, standard addition, isotopic dilution)... [Pg.201]


See other pages where Gradient Calibration is mentioned: [Pg.75]    [Pg.219]    [Pg.219]    [Pg.224]    [Pg.260]    [Pg.210]    [Pg.210]    [Pg.212]    [Pg.214]    [Pg.177]    [Pg.49]    [Pg.49]    [Pg.52]    [Pg.341]    [Pg.90]    [Pg.322]    [Pg.4986]    [Pg.439]    [Pg.497]    [Pg.501]    [Pg.348]    [Pg.283]    [Pg.283]    [Pg.579]   


SEARCH



Calibrations pulsed field gradients

Gradient techniques calibration

Magnetic field gradient calibration

© 2024 chempedia.info