Evaluation of Medical Devices in Europe

As a general rule, clinical data are required as evidence to support conformity with the requirements of the Active Implantable Medical Devices (AIMD) and the Medical Device (MD) directives with regards to safety and effectiveness under the normal conditions of use, evaluation of undesirable side effects, and the acceptability of the benefit/risk ratio. Risk analysis should be used to establish key objectives that need to be addressed by clinical data, or alternatively to justify why clinical data are not required (mainly for Class I devices). The risk analysis process should help the manufacturer to identify known (or reasonably foreseeable) hazards associated with the use of the device, and decide how best to investigate and estimate the risks associated with each hazard. The clinical data should then be used to establish the safety and effectiveness of the device under the intended use conditions, and to demonstrate that any of the residual risks are acceptable, when weighed against the benefits derived from use of the device. [Pg.187]

The provision of clinical data in support of conformity with the requirements of the directives does not necessarily mean that the device has to be subjected to actual clinical investigations. Instead, the manufacturer may present critical evaluations of [Pg.187]