Eluant generator

A much more elegant way to prepare hydroxide eluants has recently been introduced by liu et al. [148]. The respective device was commercialized under the trade name Eluant Generator. The most important part of this module is a cartridge, in which potassium hydroxide is generated by means of electrolysis. As schematically depicted in Fig. 3-146, such a cartridge consists of an electrolyte reservoir, which is filled with KOH solution (c = 4 mol/L) and connected to an electrolysis chamber via a stack of cation exchange membranes. [Pg.196]

KOH (with Eluant Generator ) gradient 9 mmol/L isocratically for 7 min, then to 46 mmol/L in 11 min flow rate 1.6 mL/min detection suppressed conductivity injection volume 2000 pL solute concentrations 2 pg/L fluoride (1),... [Pg.200]

Fig. 9-16. Separation of haloacetic acids on a hydroxide-selective anion exchanger using a gradient elution technique. - Separator column lonPac AS16 eluant KOH with Eluant Generator gradient linear, 3 mmol/L to 60 mmol/L in 25 min flow rate 1 mL/min detection suppressed conductivity injection volume 25 pL solute concentrations (a)... Fig. 9-16. Separation of haloacetic acids on a hydroxide-selective anion exchanger using a gradient elution technique. - Separator column lonPac AS16 eluant KOH with Eluant Generator gradient linear, 3 mmol/L to 60 mmol/L in 25 min flow rate 1 mL/min detection suppressed conductivity injection volume 25 pL solute concentrations (a)...
Fig. 9-39. Trace anion analysis with lonPac AS15 via direct injection utilizing an Eluant Generator. — Separator column lonPac AS15 (2-mm) eluant KOH gradient 8 mmol/L for 6 min isocratic, then linearly to 60 mmol/L in 10 min flow rate 0.5 mL/ min detection suppressed conductivity injection volume ... Fig. 9-39. Trace anion analysis with lonPac AS15 via direct injection utilizing an Eluant Generator. — Separator column lonPac AS15 (2-mm) eluant KOH gradient 8 mmol/L for 6 min isocratic, then linearly to 60 mmol/L in 10 min flow rate 0.5 mL/ min detection suppressed conductivity injection volume ...
Y. Liu, N. Avdalovic, J. Rohrer, M. Laikht-man, E. Kaiser and H, Dhillon, Determination of Trace Level Anions in High Purity Water by Ion Chromatography with On-line Eluant Generation. Presentation at the International Ion Chromatography Symposium, Osaka, Japan 1998. [Pg.860]


See other pages where Eluant generator is mentioned: [Pg.1]    [Pg.198]    [Pg.198]    [Pg.199]    [Pg.206]    [Pg.223]    [Pg.295]    [Pg.591]    [Pg.627]    [Pg.628]    [Pg.637]    [Pg.638]    [Pg.650]    [Pg.750]    [Pg.750]    [Pg.751]    [Pg.814]    [Pg.817]    [Pg.836]    [Pg.837]   
See also in sourсe #XX -- [ Pg.2 , Pg.2 , Pg.2 , Pg.2 , Pg.196 , Pg.206 , Pg.223 , Pg.295 , Pg.591 , Pg.627 , Pg.750 , Pg.814 ]