Column Dimensions and Film Thickness with Parameters in the Fundamental Resolution Equation

Correlation of Column Dimensions and Film Thickness with Parameters in the Fundamental Resolution Equation [Pg.167]

Column inner diameter affects not only resolution but also sample capacity. For example, the megabore column has an excellent capacity at the expense of a sacrifice in resolution offered by the narrower-diameter columns. Analysis times are also shorter, and of course, it remains the choice of choice for direct replacement of a packed column. But it is a column of choice for purity or trace component [Pg.168]

Thick films are necessary for separations of low boilers, thin-film columns for high boilers. As the column i.d. increases, thicker films are often selected thicker films of stationary phase also mean more column bleed (Section 3.8). One must [Pg.169]

Let us now review the fundamental resolution expression [equation (2.13)], which is repeated here for convenience, and examine each term on the right-hand side  [Pg.170]

Efficiency, N, is a function of the carrier gas (i.e., helium vs. hydrogen vs. nitrogen, and column length and diameter. Doubling column length yields a 41% in resolution quadrupling of colunm length doubles resolution but at the expense of a fourfold increase in both time of analysis and cost. The question to be asked Is it worth it  [Pg.170]




SEARCH



And film thickness

And resolution

Column dimensions

Column film thickness

Columns fundamentals

Dimensions fundamental

Films resolution

Resolution column

Resolution dimensions

Resolution equation

Resolution fundamental equation

Resolution parameter

Resolution with

The Resolution Equation

The fundamental equations

The parameters

Thick film columns

Thick films

Thick-film and

Thickness parameters

© 2024 chempedia.info