Cobyrinate nitrite-selective electrodes

Although the ISEs based on cobyrinates have good selectivity for nitrite over several anions, they also respond to salicylate and thiocyanate. To eliminate this interference, the nitrite-selective electrode based on ionophore 2 was placed behind a microporous gas-permeable membrane (GPM) in a nitrogen oxide gas-sensor mode (75). NOx was generated from nitrite in the sample at pH 1.7 and, after crossing the GPM, was trapped as nitrite by an internal solution that was buffered at pH 5.5 (0.100 M MES-NaOH, pH 5.5, containing 0.100 M NaCl). The internal solution was "sandwiched" between the nitrite-selective electrode and the GPM. [Pg.185]

Stepanek R, Krautler B, Schulthess P, Lindemann B, Ammann D, Simon W (1986) Aquocyanocobalt(ni)-hepta(2-phenylethyl)-cobyrinate as a cationic carrier for nitrite-selective liquid-membrane electrodes. Anal Chim Acta 182 83-90... [Pg.237]

A hydrophobic cobyrinate (Figure 2, structure 2) was used to prepare solvent polymeric membranes (10). The typical membrane composition was 1% (w/w) ionophore, 66% (w/w) plasticizer and 33% (w/w) polymer. Electrodes prepared with this ionophore, dioctyl sebacate (DOS) and poly(vinyl chloride) (PVC) presented, at pH 6.6, the selectivity pattern shown in Figure 3. The response of the electrodes was near-Nernstian for salicylate, thiocyanate, and nitrite. Their selectivity behavior clearly deviates from that of the Hofmeister series, with nitrite being the anion that presents the larger deviation. [Pg.181]

In dicyanocobalt(III) a,b,c,d,e,g-hexamethyl-f-stearylamide cobyrinate (derivative 3) the six peripheral amide groups of vitamin B12 have been replaced with methyl ester groups, and the proximal base of the vitamin at the f-position with a stearylamide group (11). Electrodes prepared with this ionophore and DOS as the plasticizer were also selective for thiocyanate and nitrite over the rest of the anions tested. The main anionic interferent was salicylate. In all cases, the response of the electrodes to the preferred anions was sub-Nemstian. Overall, the selectivity pattern obtained with ionophore 3 is similar to that of 2 and to that of the hydrophobic cobyrinate-based electrodes reported previously (5, 12, 13). This observation suggests that in all cobyrinate ionophores the anions interact with the cobalt(III) center, and that the side chains of the corrin ring have a small effect on the selectivity of this interaction. [Pg.181]


See also in sourсe #XX -- [ Pg.16 ]




SEARCH



Cobyrinates

Nitrite electrodes

© 2024 chempedia.info