Axial TOF Mass Spectrometers

In axial TOF instruments all ions of charge q are first accelerated by a potential difference U, between the sample support and a nearby grid or open electrode, to [Pg.53]

for a given experiment the calibration equation is m/z = At, where A is the calibration constant which is determined by least-squares fitting of this equation to a series of peaks, the m/z values of which are known and the arrival times measured. However, in order to account for an electronic delay of the trigger signal [Pg.54]


See other pages where Axial TOF Mass Spectrometers is mentioned: [Pg.53]   