Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Analytical signal generation

To maximize analytical signal generation, the ideal receptor will therefore be chosen or designed by consideration of the electrostatics of the binding site, the location of membrane positioning, intra-membrane protrusion/anchorage, association with adjacent lipids and conformational alteration. [Pg.360]

Englebienne P (2000) Analytical signal Generation and detection. Englebienne P Immune and Receptor Assays in Theory and Practice, pp. 227-289. Boca Raton, EL CRC Press. [Pg.3466]

The electron temperature is a measure of the kinetic energy of the electrons. It is relevant in the study of excitation and ionization by collisions with electrons, which is an important process for analyte signal generation. The electron temperature can be determined from the intensity of the recombination continuum or of the bremsstrahlung (Eq. 62). [Pg.639]

Covalent immobilization methods are the most difficult to employ, but this disadvantage is offset by the control afforded over packing density and strand orientation. The use of covalent attachment of oligomers has been observed to provide a very stable means of oligonucleotide attachment and, in conjunction with substrate linker molecules of sufficient length (>25 atoms) have demonstrated fast kinetics of hybridization where analytical signal generation was observed to occur in minutes as opposed to hours. In order to identify the surface derivatisation conditions which provide the optimal sensor response characteristics such as response time. [Pg.284]

The formation of PPy on an electrode surface provides a nanoporous matrix that is highly used for the immobilization of biomolecules to design various biosensors (electrochemical biosensor, immunosensor, and DNA sensor). It also acts as a mediator to transfer the analytical signal generated by some redox enzymes to the transducer even if the redox center is deeply buried in the protein globule. In addition, it is an efficient protector of electrodes against interfacing materials (proteins present in real samples such as blood and urine). [Pg.77]

A standardization is still possible if the analyte s signal is referenced to a signal generated by another species that has been added at a fixed concentration to all samples and standards. The added species, which must be different from the analyte, is called an internal standard. [Pg.116]

A single-channel manifold also can be used for systems in which a chemical reaction generates the species responsible for the analytical signal. In this case the carrier stream both transports the sample to the detector and reacts with the sample. Because the sample must mix with the carrier stream, flow rates are lower than when no chemical reaction is involved. One example is the determination of chloride in water, which is based on the following sequence of reactions. ... [Pg.652]

Optimization lefeis to the step in the analytical process (Fig. 2) where some sort of treatment is performed on samples to generate taw data which can be in the form of voltages, currents, or other analytical signals. These data have yet to be caUbrated in terms of chemical concentrations. [Pg.430]

Quantitation using mass spectrometry is no different to quantitation using other techniques and, as discussed above in Section 2.5, involves the comparison of the intensity of a signal generated by an analyte in a sample to be determined with that obtained from standards containing known amounts/concentrations of that analyte. [Pg.70]

Figure 2. Signals generated in a thin specimen by a focused electron beam in an analytical electron microscope. Figure 2. Signals generated in a thin specimen by a focused electron beam in an analytical electron microscope.
The analytical response generated by an immunoassay is caused by the interaction of the analyte with the antibody. Although immunoassays have greater specificity than many other analytical procedures, they are also subject to significant interference problems. Interference is defined as any alteration in the assay signal different from the signal produced by the assay under standard conditions. Specific (cross-reactivity) and nonspecific (matrix) interferences may be major sources of immunoassay error and should be controlled to the greatest extent possible. Because of their different impacts on analyses, different approaches to minimize matrix effects and antibody cross-reactivity will be discussed separately. [Pg.683]

Several authors [386,387] have discussed the spectroscopic and nonspectroscopic (matrix) interferences in ICP-MS. ICP-MS is more susceptible to nonspectroscopic matrix interferences than ICP-AES [388-390]. Matrix interferences are perceptible by suppression and (sometimes) enhancement of the analyte signal. This enhanced susceptibility has to be related to the use of the mass spectrometer as a detection system. In fact, since both techniques use the same (or comparable) sample introduction systems (nebulisers, spray chambers, etc.) and argon plasmas (torches, generators, etc.), it is reasonable to assume that, as far as these parts are concerned, interferences are comparable. The most severe limitation of ICP-MS consists of polyatomic... [Pg.655]

In the following, the stages of the analytical process will be dealt with in some detail, viz. sampling principles, sample preparation, principles of analytical measurement, and analytical evaluation. Because of their significance, the stages signal generation, calibration, statistical evaluation, and data interpretation will be treated in separate chapters. [Pg.42]

Analytical signals are generated by interactions between species of the analyte, to be precise between certain forms of intrinsic energy of them (see... [Pg.70]

The generation of analytical signals is a complex process that takes place in several steps. Methods of instrumental analysis often need five steps, namely (1) the genesis, (2) the appearance, (3) the detection and conversion, (4) the registration, and (5) the presentation of signals see Fig. 3.3. [Pg.72]

Fig. 33. Illustration of the steps of signal generation for different analytical principles... Fig. 33. Illustration of the steps of signal generation for different analytical principles...
In principle, there is no fundamental contrast between qualitative, (semi-quantitative), and quantitative analyses. The analytical signal is generated in the same way, only the detection and evaluation is done on the basis of a more rough scale, in qualitative analysis only in form of a yes/no decision. [Pg.111]

More than 90% of commercially available enzyme-based biosensors and analytical kits contain oxidases as terminal enzymes responsible for generation of analytical signal. These enzymes catalyze oxidation of specific analyte with molecular oxygen producing hydrogen peroxide according to the reaction ... [Pg.448]

Although x-ray microanalysis in the STEM is the most developed form of analytical electron microscopy, many other types of information can be obtained when an electron beam interacts with a thin specimen. Figure 2 shows the various signals generated as electrons traverse a thin specimen. The following information about heterogeneous catalysts can be obtained from these signals ... [Pg.307]

The intrinsic sensors are based on the direct recognition of the chemicals by its intrinsic optical activity, such as absorption or fluorescence in the UV/Vis/IR region. In these cases, no extra chemical is needed to generate the analytical signal. The detection can be a traditional spectrometer or coupled with fiber optics in those regions. Sensors have been developed for the detection of CO, C02 NOx, S02, H2S, NH3, non-saturated hydrocarbons, as well as solvent vapors in air using IR or NIR absorptions, or for the detection of indicator concentrations in the UV/ Vis region and fluorophores such as quinine, fluorescein, etc. [Pg.761]

The separation of binding site and fluorophore by a nonconjugating spacer opens the path to other mechanisms of communication, most prominently ET and exci-mer/exciplex formation. In the first case, the electronic nature of both fluorophore and receptor unit and the steric nature of the spacer are the important parameters for signal generation. In the second case, for most systems the electronic nature of the fluorophores and the steric nature of the receptor as well as its change upon analyte binding determine the signal. [Pg.50]


See other pages where Analytical signal generation is mentioned: [Pg.355]    [Pg.355]    [Pg.68]    [Pg.395]    [Pg.194]    [Pg.550]    [Pg.176]    [Pg.44]    [Pg.670]    [Pg.683]    [Pg.313]    [Pg.670]    [Pg.249]    [Pg.41]    [Pg.94]    [Pg.57]    [Pg.155]    [Pg.157]    [Pg.441]    [Pg.244]    [Pg.324]    [Pg.5]    [Pg.29]    [Pg.132]    [Pg.755]    [Pg.151]    [Pg.515]    [Pg.28]    [Pg.43]   
See also in sourсe #XX -- [ Pg.46 ]

See also in sourсe #XX -- [ Pg.46 ]




SEARCH



Analyte signal

Analytic signal

Analytical signals

Signal generation

Signal generator

Signaling/signal generation

© 2024 chempedia.info