Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Second difference waveform

It has been shown in Section 4.1 that replacing sinusoidal modulation by a digital form makes for a simpler expression for the spectral signal, and that the modulation amplitude can be selected to maximise its intensity. In fact the waveform demonstrated in Figure 4.2B, synchronised to waveform A, s5mthesises the second difference of the spectral function ... [Pg.68]

The predictable nature of crack advance under these conditions admits two simple consequences for cyclic loading scenarios. First, for a given waveform, the number of cycles should be irrelevant because the time integral of the power law is insensitive to frequency. Second, for two different waveforms (e.g., sine and square) of the same stress range and load ratio, the time integral of the FCP power law will not be identical, and thus some waveforms will appear more detrimental than others in a mathematically predictable manner. Fuimanski and Pruitt tested these two consequences of static mode FCP behavior in conventional UHMWPE to evaluate the extent of the applicability of the static mode model [58]. [Pg.457]

Waveform can be defined in at least two different ways which are relevant to spectroscopic measurements. Wavelength (X.) is defined as the distance between successive peaks (Figure 2.1) and is measured in subunits of a metre, of which the most frequently used is the nanometre (10 9m). An angstrom unit (A) is not acceptable in SI terminology but is still occasionally encountered and is 10-10m (i.e. 10 A = 1 nm). The frequency of radiation (nu, v) is defined as the number of successive peaks passing a given point in 1 second. Hence the relationship between these two units of measurement is ... [Pg.37]

Fig. 13. Diagram to show the effect of a rising baseline during the recovery of a fluorescence waveform with a boxcar integrator, (a) shows the results with a single scanning aperture and (b) how a second fixed aperture can be used to sample the baseline. The output curve is the difference between the signal in the scanning aperture and that sampling the baseline. Fig. 13. Diagram to show the effect of a rising baseline during the recovery of a fluorescence waveform with a boxcar integrator, (a) shows the results with a single scanning aperture and (b) how a second fixed aperture can be used to sample the baseline. The output curve is the difference between the signal in the scanning aperture and that sampling the baseline.
Figure 11.15 shows how the electron clouds for the electrons of the atoms of first and second period elements are envisioned. Note that the electron charge waveforms for the electrons in different orbitals of each atom are all superimposed on each other. [Pg.427]

The reflection wave form of limestone is W shape and the second wave trough is shallow, this indicates that the limestone specimens has not badly damaged. Moreover, the transmitted wave does not descend immediately at maximum stress, but lasts a period of time, which shows that limestone specimens did not serious damage like the anthracite. From the waveform curve anthracite coal and limestone, there are significant differences between the two materials in terms of dynamic mechanical properties. [Pg.53]

Schwan (1957) wrote an introduction to impedance measurement on cell and bacteria suspensions. A Coulter counter is a streaming blood cell counter. Two different electrolyte reservoirs equipped with impedance CC electrodes are connected via a short capillary. The reservoir with the cell suspension is on higher pressure so that the electrolyte with cells stream through the capillary. Measured impedance is dominated by the capillary because of its very small diameter. At passage, the impedance goes through a peak because of the cell membrane. The frequency of the peaks is the number of cells passed per second the impedance waveforms are characteristic for different cell types. The cell suspension must have a concentration so that the probability of two cells in the capillary at the same time is low. The electrolyte concentration must be adapted to the impedance increase caused by each cell. [Pg.177]


See other pages where Second difference waveform is mentioned: [Pg.69]    [Pg.69]    [Pg.239]    [Pg.321]    [Pg.350]    [Pg.331]    [Pg.335]    [Pg.260]    [Pg.155]    [Pg.158]    [Pg.161]    [Pg.321]    [Pg.158]    [Pg.165]    [Pg.185]    [Pg.333]    [Pg.25]    [Pg.24]    [Pg.154]    [Pg.7]    [Pg.173]    [Pg.87]    [Pg.6465]    [Pg.51]    [Pg.110]    [Pg.420]    [Pg.260]    [Pg.322]    [Pg.473]    [Pg.287]    [Pg.6]    [Pg.6464]    [Pg.374]    [Pg.135]    [Pg.524]    [Pg.331]    [Pg.254]    [Pg.478]    [Pg.95]    [Pg.191]    [Pg.276]    [Pg.471]   
See also in sourсe #XX -- [ Pg.68 , Pg.69 ]




SEARCH



Second difference

© 2024 chempedia.info