Procedure A Case Study

Since calibration is time consuming and is often viewed as nonproductive work, the decisions in developing a calibration program often [Pg.293]

TABLE I Summary of Calibration Tests and Acceptance Criteria [Pg.294]

HPLC module Test Procedure Acceptance criteria [Pg.294]

Pump Flow accuracy Run pump at 0.3 and 1.5 mL/min (65% acetonitrile/water) and collect 5 mL from detector into a volumetric flask. Measure time. 5% [Pg.294]

Column Temperature Check actual column oven 35 2°C [Pg.294]


See other pages where Procedure A Case Study is mentioned: [Pg.291]    [Pg.293]    [Pg.128]   