Plate height programmed temperature

Helium is the most common carrier gas and is compatible with most detectors. For a flame ionization detector, N2 gives a lower detection limit than He. Figure 24-11 shows that H2, He, and N2 give essentially the same optimal plate height (0.3 mm) at significantly different flow rates. Optimal flow rate increases in the order N2 < He < H2. Fastest separations can be achieved with H, as carrier gas, and H2 can be run much faster than its optimal velocity with little penalty in resolution.11 Figure 24-12 shows the effect of carrier gas on the separation of two compounds on the same column with the same temperature program. [Pg.537]