The concept of the triple quadmpole instrument may be expanded by adding an-otha- RF-only quadmpole and a third mass-analyzing quadmpole to build a QqQqQ instrument. The penta quadmpole represents one of the rare tandem-inspace concepts to achieve MS. A commercial penta quadmpole instmment has once been offered by Extrel, but remained fairly exotic. Nonetheless, the penta quadmpole instmment may serve as a versatile tool for the study of gas phase ion chemistry and in particular ion-molecule reactions [99-102]. For the latter application the effective spatial separation of the zone with neutral reagent from that of preparation of the /n/z-selected precursor ion presents a clear advantage over tandem-in-time concepts. [Pg.438]

Product ion, define m-i no scan, select m-i metastable or CID scan up to to collect its fragments [Pg.439]

Precursor ion, define m2 scan from m2 upwards to cover potential precursors metastable or CID no scan, select m2 [Pg.439]

Constant neutral loss, define n scan desired range metastable or CID scan range shifted by Am to low mass [Pg.439]


See other pages where Penta Quadrupole Instruments is mentioned: [Pg.438]    [Pg.438]    [Pg.2881]   