PANSY-TOCSY

Fig. 5 PANSY-TOCSY pulse sequence for simultaneous acquisition of two-dimensional H-C HETCOR and H-H TOCSY spectra... Fig. 5 PANSY-TOCSY pulse sequence for simultaneous acquisition of two-dimensional H-C HETCOR and H-H TOCSY spectra...
J-splittings are only present in the C-13 detected spectra. Not only do J-splittings reduce the intensity of the correlation peaks and make the spectra more crowded and difficult to analyse, but such splittings often introduce severe overlap problems. Therefore decoupling A and X spins is generally desirable. This is easily achieved by displacing the A and X acquisition periods. The 2D H-C PANSY TOCSY experiment shown in Fig. 5 does just that. [Pg.77]

Fig. 6 Two-dimtaisi Fig. 6 Two-dimtaisi<Hial PANSY-TOCSY spectra of brucine recraded in ca. 20 min on a Varian 600 MHz NMR system equipped with two receivers, (a) H-C HETCOR spectrum and (b) H-H TOCSY spectrum 122 ms Iraig mixing period, 128 cmnplex increments and 2 scans ptr increment were used. Reproduced limn Ref. 27. The Ammican Chcanical Society, 2006...
An example of the Hadamard encoded PANSY-TOCSY pulse sequence is shown in Fig. 18. The conventional free evolution is replaced with a Hadamard encoding pulse as can be appreciated by comparing Figs. 17 and 5. The PANSY-TOCSY spectra of inosine recorded with Hadamard encoding in just 22 s are shown in Fig. 19. [Pg.90]

Fig. 18 PANSY-TOCSY pulse sequence with Hadamard encoding (to be compared with the conventional pulse sequence shown in Fig. 5)... Fig. 18 PANSY-TOCSY pulse sequence with Hadamard encoding (to be compared with the conventional pulse sequence shown in Fig. 5)...
Fig. 19 Hadamard PANSY-TOCSY spectra of inosine recorded in 22 s using Hadamard encoding. A Varian 600 MHz system equipped with two receivers and a cryogenic HCN probe that is optimized for C-13 detection were used. The pulse sequence is shown in Fig. 17. Reproduced from Ref. 27. The American Chemical Society, 2006... Fig. 19 Hadamard PANSY-TOCSY spectra of inosine recorded in 22 s using Hadamard encoding. A Varian 600 MHz system equipped with two receivers and a cryogenic HCN probe that is optimized for C-13 detection were used. The pulse sequence is shown in Fig. 17. Reproduced from Ref. 27. The American Chemical Society, 2006...

See other pages where PANSY-TOCSY is mentioned: [Pg.78]    [Pg.78]    [Pg.78]    [Pg.78]    [Pg.71]   
See also in sourсe #XX -- [ Pg.77 ]