Multi hyphenated GC techniques

Gas chromatography is a most favourable case for interfacing to a mass spectrometer, as the mobile phases commonly used do not generally influence the spectra observed, and the sample, being in the vapour phase, is compatible with the widest range of mass-spectral ionisation techniques. The primary incompatibility in the case of GC-MS is the difference in operating pressure for the two hyphenated instruments. The column outlet in GC is typically at atmospheric pressure, while source pressures in the mass spectrometer range from 2 to [Pg.456]

GC-AAS has found late acceptance because of the relatively low sensitivity of the flame graphite furnaces have also been proposed as detectors. The quartz tube atomiser (QTA) [186], in particular the version heated with a hydrogen-oxygen flame (QF), is particularly effective [187] and is used nowadays almost exclusively for GC-AAS. The major problem associated with coupling of GC with AAS is the limited volume of measurement solution that can be injected on to the column (about 100 xL). Virtually no GC-AAS applications have been reported. As for GC-plasma source techniques for element-selective detection, GC-ICP-MS and GC-MIP-AES dominate for organometallic analysis and are complementary to PDA, FTIR and MS analysis for structural elucidation of unknowns. Only a few industrial laboratories are active in this field for the purpose of polymer/additive analysis. GC-AES is generally the most helpful for the identification of additives on the basis of elemental detection, but applications are limited mainly to tin compounds as PVC stabilisers. [Pg.456]

Finally, we can note that GC or GC-MS in general are capable of analysing only 20 % of the chemicals in use worldwide. [Pg.456]

Principles and Characteristics Gas chromatogra-phy/infrared spectrometry systems couple the separation capability of GC with the selectivity for functional [Pg.456]

The success of spectral identification depends on the appropriate reference spectra for comparison. IR measurement of eluates that are at slightly subambient temperature is advantageous considering that the large databases of condensed-state spectra may be searched. Spectra measured by matrix-isolation GC-FTIR have characteristically narrow bandwidths compared with the spectra of samples in the condensed phase near ambient temperature or in the gas phase. In addition, the relative intensities of bands in the spectra of matrix-isolated samples often change compared with either gas- or condensed-phase spectra [195]. GC-FTIR spectra obtained by direct deposition match well with the corresponding reference spectra in standard phase [Pg.457]


See other pages where Multi hyphenated GC techniques is mentioned: [Pg.425]    [Pg.456]   