Measurement of the Stokes parameters

The electric field (in Jones vector form) of the incident light (before the quarter-wave plate) is [Pg.100]

At the polarizer, the electric field along the transmission axis is [Pg.101]

In the first step of the measurement, the quarter-wave plate is removed, which is equivalent to P = Q. The light intensity is measured when the polarizer is at the following positions  [Pg.101]

When the quarter-wave plate is inserted and the polarizer is at 45°, the measured light intensity will be [Pg.101]

From these four equations, the Stokes parameters can be calculated. [Pg.101]


Figure 19. The experimental setup for measurement of the Stokes parameters. Figure 19. The experimental setup for measurement of the Stokes parameters.
Figure 11.4 Measurement of the Stokes parameters with a retarder and an ideal perfect linear polarizer rotated with respect to the laboratory reference frame. Figure 11.4 Measurement of the Stokes parameters with a retarder and an ideal perfect linear polarizer rotated with respect to the laboratory reference frame.