Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Mass quadrupole instruments

An example of linked scanning on a triple quadrupole instrument. A normal ion spectrum of all the ions in the ion source is obtained with no collision gas in Q2 all ions scanned by Q1 are simultaneously scanned by Q3 to give a total mass spectrum (a). With a collision gas in Q2 and with Q1 set to pass only m+ ions in this example, fragment ions (f, fj ) are produced and detected by Q3 to give the spectrum (b). This CID spectrum indicates that both f, and fj are formed directly from m+. [Pg.234]

Most ion sources produce singly charged ions, i.e., z = I and the ranges shown here apply to such ions. Matrix assisted methods may produce ions with r > 1. When = 1, m/z. = m, viz., mass can be measured directly. An ES ion source produces ions with z > 1 and this effectively extends the mass ranges that can be examined. For example, with z = 1 and m = 10,000, the m/z value is 10,000 and this would be beyond tbe capabilities of a quadrupole instrument. [Pg.282]

Electrospray can be used with sector, time-of-flight, and quadrupole instruments. The technique has been used extensively to couple liquid chromatographs to mass spectrometers. [Pg.390]

By using a sampling device, the ions are siphoned from the end of the plasma flame and led into an ion mass analyzer, such as a quadrupole instrument, where the abundances of the ions and their m/z values are recorded. [Pg.395]

A discharge ignited in argon and coupled inductively to an external high-frequency electromagnetic field produces a plasma of ions, neutrals, and electrons with a temperature of about 7000 to 10,000°C. Samples introduced into the plasma under these extremely energetic conditions are fragmented into atoms and ions of their constituent elements. These ions are examined by a mass analyzer, frequently a quadrupole instrument. [Pg.395]

Mass spectrometer configuration. Multianalyzer instruments should be named for the analyzers in the sequence in which they are traversed by the ion beam, where B is a magnetic analyzer, E is an electrostatic analyzer, Q is a quadrupole analyzer, TOP is a time-of-flight analyzer, and ICR is an ion cyclotron resonance analyzer. For example BE mass spectrometer (reversed-geometry double-focusing instrument), BQ mass spectrometer (hybrid sector and quadrupole instrument), EBQ (double-focusing instrument followed by a quadrupole). [Pg.430]

SiiH2 are completely separated from the peak. Quadrupole instruments are not usually capable of such high mass resolution. [Pg.544]

The choice of mass spectrometer for a particular analysis depends on the namre of the sample and the desired results. For low detection limits, high mass resolution, or stigmatic imaging, a magnetic sector-based instrument should be used. The analysis of dielectric materials (in many cases) or a need for ultrahigh depth resolution requires the use of a quadrupole instrument. [Pg.548]

The triple quadrupole instrument consists of two mass analyzers separated by an rf-only quadrupole. In the rf-only mode, ions of all masses are... [Pg.14]

Resolution Quadrupole instruments are not capable of achieving the high resolution that is common with double-focusing magnetic-sector instruments. In GC/MS analyses, a compromise is struck between sensitivity (ion transmission) and mass resolution. In the quadrupole instrument, the resolution is set to the lowest possible value commensurate with resolving peaks differing by 1 Dalton (unit resolution). [Pg.204]

The most important gas phase analytical techniques are mass spectrometry and gas phase chromatography. If the total gas pressure is <10-4-10-8 Torr, a mass spectrometer such an omegatron or a quadrupole instrument may be inserted into the reactant volume. However, in most cases, the pressure is in excess of this, and gas must be delivered to a mass spectrometer via a leak, such as a Metrosil pellet or a capillary constriction, situated as closely as possible to the reaction volume. [Pg.18]

The layout of an ICP-MS is shown schematically in Figure 8.17 and comprises three essential parts the ICP torch, the interface and the mass spectrometer. The ICP torch differs little from that discussed earlier and the mass spectrometer is very similar to those used for organic mass spectrometry and discussed in Chapter 9. Typically a quadrupole instrument would be used. The construction of the interface is shown in Figure 8.18 and is based on the use of a pair of water-cooled cones which divert a portion of the sample stream into the ion optics of the mass spectrometer whence the mass spectrum is produced by standard mass spectrometer operation. Some modern instruments also incorporate a so-... [Pg.308]

Fig. 11.16. Representation of three tandem mass spectrometry (MS/MS) scan modes illustrated for a triple quadrupole instrument configuration. The top panel shows the attributes of the popular and prevalent product ion CID experiment. The first mass filter is held at a constant m/z value transmitting only ions of a single mlz value into the collision region. Conversion of a portion of translational energy into internal energy in the collision event results in excitation of the mass-selected ions, followed by unimolecular dissociation. The spectrum of product ions is recorded by scanning the second mass filter (commonly referred to as Q3 ). The center panel illustrates the precursor ion CID experiment. Ions of all mlz values are transmitted sequentially into the collision region as the first analyzer (Ql) is scanned. Only dissociation processes that generate product ions of a specific mlz ratio are transmitted by Q3 to the detector. The lower panel shows the constant neutral loss CID experiment. Both mass analyzers are scanned simultaneously, at the same rate, and at a constant mlz offset. The mlz offset is selected on the basis of known neutral elimination products (e.g., H20, NH3, CH3COOH, etc.) that may be particularly diagnostic of one or more compound classes that may be present in a sample mixture. The utility of the two compound class-specific scans (precursor ion and neutral loss) is illustrated in Fig. 11.17. Fig. 11.16. Representation of three tandem mass spectrometry (MS/MS) scan modes illustrated for a triple quadrupole instrument configuration. The top panel shows the attributes of the popular and prevalent product ion CID experiment. The first mass filter is held at a constant m/z value transmitting only ions of a single mlz value into the collision region. Conversion of a portion of translational energy into internal energy in the collision event results in excitation of the mass-selected ions, followed by unimolecular dissociation. The spectrum of product ions is recorded by scanning the second mass filter (commonly referred to as Q3 ). The center panel illustrates the precursor ion CID experiment. Ions of all mlz values are transmitted sequentially into the collision region as the first analyzer (Ql) is scanned. Only dissociation processes that generate product ions of a specific mlz ratio are transmitted by Q3 to the detector. The lower panel shows the constant neutral loss CID experiment. Both mass analyzers are scanned simultaneously, at the same rate, and at a constant mlz offset. The mlz offset is selected on the basis of known neutral elimination products (e.g., H20, NH3, CH3COOH, etc.) that may be particularly diagnostic of one or more compound classes that may be present in a sample mixture. The utility of the two compound class-specific scans (precursor ion and neutral loss) is illustrated in Fig. 11.17.
This basically means that two instruments have been linked together. The first analyser can replace the traditional chromatographic separation step and is used to produce ions of chosen m/z values. Each of the selected ions is then fragmented by collision with a gas, and mass analysis of these product ions effected in the second analyser. The resulting mass spectrum is used for their identification. The potential combinations of the various magnetic sector and quadrupole instruments to form such coupled systems is considerable. Ion traps may also be operated in a tandem MS mode. [Pg.128]


See other pages where Mass quadrupole instruments is mentioned: [Pg.185]    [Pg.227]    [Pg.231]    [Pg.278]    [Pg.14]    [Pg.14]    [Pg.1029]    [Pg.494]    [Pg.314]    [Pg.444]    [Pg.767]    [Pg.496]    [Pg.386]    [Pg.389]    [Pg.400]    [Pg.401]    [Pg.402]    [Pg.406]    [Pg.106]    [Pg.57]    [Pg.334]    [Pg.334]    [Pg.64]    [Pg.51]    [Pg.51]    [Pg.231]    [Pg.16]    [Pg.332]    [Pg.383]    [Pg.384]    [Pg.719]    [Pg.195]    [Pg.199]    [Pg.400]    [Pg.288]    [Pg.55]   
See also in sourсe #XX -- [ Pg.111 ]




SEARCH



Instrumentation quadrupole

Instrumentation quadrupoles

Mass quadrupole

© 2024 chempedia.info