Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Intensity limiting molar mass

Figures 5.24, 5.25 and 5.26 also show that the limiting molar masses are lower the higher the intensity. Whilst Okuyama [50] and Thomas et al. [51] predicted, and several workers observed [52], that the limiting molar mass is invariant with intensity, most workers now agree that decreases with increase in ultrasonic intensity. Price [39] found that the results of the ultrasonic degradation of polystyrene in toluene fitted equation (Eq. 5.22). Figures 5.24, 5.25 and 5.26 also show that the limiting molar masses are lower the higher the intensity. Whilst Okuyama [50] and Thomas et al. [51] predicted, and several workers observed [52], that the limiting molar mass is invariant with intensity, most workers now agree that decreases with increase in ultrasonic intensity. Price [39] found that the results of the ultrasonic degradation of polystyrene in toluene fitted equation (Eq. 5.22).
This variation in limiting molar mass vith intensity can be predicted if we accept Schmid s hypothesis that a modified Stake s equation can be applied to polymer degradation. [Pg.182]

Tab. 5.19 Limiting molar mass values as a function of temperature (intensity = 25 Wcm ... Tab. 5.19 Limiting molar mass values as a function of temperature (intensity = 25 Wcm ...
The thermal lability of the R-C-O-N bond system controls the reversibility of the chain termination and limits also the use of NMP. SFRP of styrene at about 130 °C is studied intensively. In this case, high control and high-molar-mass products could be achieved. It was found that the thermal autopolymerization... [Pg.185]

As the main aim of trace analysis is usually determination of the mass (expressed as the number of moles) of a given component in a studied sample, molar concentration is generally not used. Some exceptions are electrochemical methods, where the analytical signal (e.g., current intensity) is a direct function of molar concentration [9]. Therefore, in voltamperometric techniques the detection limits are usually given in molar concentration units (Table 1.2). Thus, for nickel (molar mass M = 58.7 g/mol) the detection limit in inverse voltammetry is approximately 6 X 10 mol/L, and is expressed as a mass fraction, 3.5 x 10 g/dm, or as a percentage, 0.35 x 10 %. In spectrophotometry, when concentrations are given in molar units then molar absorptivities are also used. For example, molar absorptivity e = 5xl0 " L/mol cm corresponds (for molar mass M = 58.7 g/mol) to molar absorptivity a = 5 x 10 /(58.7 x 10 )mL/g cm (i.e., 0.85 mL/g cm). [Pg.8]

The thermal lability of the R—C—O—N bond system controls the reversibility of the chain termination and limits also the use of NMRP. SFRP of styrene at about 130°C is studied intensively. In this case, high control and high molar mass products could be achieved. It was found that the thermal autopolymerization of the styrene monomer plays an important role in the mechanism of the reaction. Therefore, first experiments using different monomers in the presence of TEMPO and a radical initiator failed with regard to the control. However, new nitroxide adducts with a different R—O—N bond stabiUty have been developed, for example, by Hawker [14], which work also for styrene derivatives as well as for acrylates. End group functionalization in NMRP can be achieved by using a functional radical initiator in combination with a stable radical or functionalized nitroxide adducts. [Pg.36]

In this case, the enthalpic interactions within the HPLC system exceed the exclusion effects (see Figure 3(e)). The retention volumes of polymer species as a rale exponentially increase with their molar masses. The important limitation of the resulting procedures was presented in section 11.5.2.3. The retention of (high) polymers is usually so intense that the latter do not elute from the column any more. Therefore, the majority of enthalpy controlled HPLC procedures is applicable only to ohgomers - up to molar mass of few thousands g.mol. Still, the reduced sample recovery may affect results of separation even in case of oligomers. The selectivity of enthalpy driven HPLC separation is much higher than in the case of SEC but, naturally, the sequence of molar masses eluted from the column is reversed. If the effect of enthalpy is reduced, problems with sample recovery are mitigated - but at the same time the separation selectivity is reduced. [Pg.304]

Even if one solves the indexing problem and then proceeds with the analysis by an evaluation of measured reflection intensities, one cannot expect to achieve an accuracy in the crystal structure data which would be comparable to those of low molar mass compounds. This is not only a result of the lack of single crystals, but represents also a principal property In small crystallites, as they are found in partially crystalline polymers, lattice constants can be affected by their size. In many cases crystallites are not only limited in chain direction by the finite thickness of the crystalline lamellae but also laterally since polymer crystallites are often composed of mosaic blocks. Existence of these blocks is indicated in electron microscopic investigations on... [Pg.156]

As discussed above, "conventional" methods for MM and MMD determination suffer from a series of limitations. It is therefore fruitful to look to other devices and techniques, such as MS, where such limitations might not be encountered. The MS method is based on the fact that ions strike the detector and produce a current that reflects the number of ions. The number of ions of mass M is proportional to the number of molecules at that mass, which implies that the detector measures the molar fraction of molecules of mass M. For polymers, one records the mass spectrum, tabulates MS peak intensities and computes Mn and Mw using Eqs. (2.1) and (2.2). Furthermore, one can directly check from the whole spectrum if the MM distribution is unimodal or bimodal. [Pg.74]

As a result, O Brien and Bowman developed a comprehensive photopolymerization model. It incorporates heat and mass transfer effects, diffusion-controlled propagation and termination, and temporal and spatial variation of species concentration, temperature, and hght intensity. This model is applied to systems with varying diermal and optical properties. The absorbance of the polymerizing system is varied by altering either the initiator concentration, sample thickness, or molar absorption coefficient of the initiator. Based on simulations they concluded that the choice of initiator and sample thickness limits the initiator concentration usable to achieve complete monomer... [Pg.180]

Semi-quantitative analysis facilitates fast and simple multi-element measurements with limited precision. ICP-MS offers excellent semi-quantitative capabilities, as a result of the high ionisation efficiencies achieved for the majority of elements and the simplicity of the resulting mass spectra. Semi-quantitative determinations are mostly based on a comparison of response tables and the actual count rates of the sample. The response (intensity I in counts/s) of an analyte ion depends on the concentration of the analyte element, the isotopic abundance of the observed isotope, the ionisation efficiency, the atomic mass and the efficiencies of nebuUsation, ion transmission and ion detection in the mass spectrometer. In most ICP-MS instraments a plot of the atomic response, Ra, versus atomic mass yields a smooth response curve, which is fitted best by a third order polynomial (Figure 4.3). The atomic response is defined by equation (4.7) and is equivalent to the molar response divided by the Avogadro constant, IVa-... [Pg.156]


See other pages where Intensity limiting molar mass is mentioned: [Pg.189]    [Pg.52]    [Pg.394]    [Pg.495]    [Pg.129]    [Pg.157]    [Pg.111]    [Pg.157]    [Pg.244]    [Pg.905]    [Pg.276]    [Pg.1198]    [Pg.2150]    [Pg.2873]    [Pg.166]    [Pg.183]    [Pg.342]    [Pg.59]    [Pg.163]    [Pg.1401]    [Pg.1418]    [Pg.743]    [Pg.291]   
See also in sourсe #XX -- [ Pg.181 ]




SEARCH



Mass intensity

Mass limit

Mass limitation

Molar limiting

Molar mass

Molar mass, limiting

Molarity molar masses

© 2024 chempedia.info