Float collision cell

The technique consists of introducing a gas into the collision cell as a molecular beam directed perpendicularly to the ion beam accelerated by a potential of a few kilovolts towards an outlet where a pump removes it from the instrument so as to preserve a vacuum compatible with the use of electromagnetic analysers. This collision cell is located at a focal point to avoid losses. It is better to float the collision cell electrically to distinguish the metastable from the CID products. [Pg.198]

Select the precursor ions by EBE mass spectrometer and introduce them into the collision cell floated at 3600 V. The kinetic energy of the ions with charge z is 400 zeV. [Pg.1114]