Enthalpy Calibration of DTA and DSC

The relationship between the peak area (A) of the DTA or DSC curve, corresponding to an [Pg.9]

Basically, there are three approaches to peak area measurement (a) the cut and weigh method, which is accurate to within 5% (b) the planimeter method, which is accurate to within 1% (c) the integrator method, which is more accurate than method (b). It should be noted that when computer-equipped apparatus is used, enthalpy calculations are carried out automatically if the apparatus is correctly calibrated using calibration reference materials. [Pg.10]

The higher the temperature in DTA measurements, the smaller is the peak area for the same heat quantity owing to heat radiation, hence the melting point of chosen calibration material [Pg.10]

Compound or element Melting point oc Enthalpy of fusion Jg  [Pg.11]

The heat of transition of a sample can be expressed as follows when it is determined by [Pg.11]


See other pages where Enthalpy Calibration of DTA and DSC is mentioned: [Pg.9]   