Enhancing the sensitivity of LC-NMR

Transfer directly into flow probe with deuterated solvent [Pg.202]

multiple trapping of a single component on the same cartridge followed by elution with a small cell matched volume ( 30 pi) of deuterated solvent to optimise the concentration in the active volume of the NMR flow cell  [Pg.203]

drying cartridges and eluting with small volumes of deuterated solvents (cost-effective) gives very little requirement for solvent suppression, leading to significantly cleaner baselines and no loss of useful resonances in the NMR spectrum  [Pg.203]

NMR sensitivity is independent of the chromatographic peak volume, allowing more efficient characterisation of late-eluting broad peaks, which were problematic by LC-NMR  [Pg.203]

good quality two-dimensional spectra (H-H and H-C) can be acquired in reasonable times for impurities at 0.3% or higher levels. For impurities at 0.2% or less cryogenic flow probe with an active volume of 30 pi is required  [Pg.203]