When not in use, HPLC columns should be kept filled with solvent, air-tight and free from vibration. Water is generally regarded as an unsuitable conservation medium in view of potential fungal growth. Salt solutions (buffers) may crystallize and cause clogging. [Pg.136]

Do not forget to label the column with the type of mobile phase. [Pg.136]

The following solvents are consecutively pumped through at a rate of 1-3 ml min  [Pg.137]

75 ml of 1-5% aqueous acetic acid if basic impurities are present  [Pg.137]

75 ml of hexane (if hexane is subsequently used for chromatography, otherwise stop at ether). [Pg.137]

If the separation performance should decrease as a result of the silica adsorbing too much water, the latter can be removed by chemical means the reagent either being self-formulated or bought as a ready-made version (Alltech Associates, Deerfield, Illinois, USA). [Pg.126]