A DETERMINATION OF CADMIUM BY FLAME AAS

1 The sample is nebulised Into an air-acetylene flame, where it is vaporised cadmium compounds are atomised and the cadmium atoms thus formed absorb radiation from a hollow cathode lamp. The absorbance is measured at a wavelength of 228.8 nm, using background correction. [Pg.56]

2 This determination may be carried out on digest 2.3 (microwave digestion with [Pg.56]

HNO3 - H2O2 - HF), digest 2.4 (digestion with HNO, - - HF), digest 2.6 [Pg.56]

1 This procedure yields a standard curve that is linear up to approximately 0.8 mg/L Cd. [Pg.56]

2 The detection limit is approximately 0.02 mg/L in the digest. The determination limit is approximately 0.06 mg/L (3.0-7.5 mg/kg in dried plant material). [Pg.56]