Shock Polarity and Configuration

The polarity in which an ICD shock is delivered is programmable in most conventional devices (initial or reversed polarity). Selecting a shock polarity comes into play when testing shock efficacy. One polarity may be successful when the other is not. Thus, when successful VF defibrillation is not accomplished to an adequate safety margin with one shock polarity a reversal of polarity is usually attempted first. This simple maneuver may yield an appropriate shock safety margin and negate the need to implant further hardware. Also, successful defibrillation is more likely to occur when the amount of left ventricular myocardium in the shock pathway is maximized. To this end some pathways and lead coiffigurations may work better than others. [Pg.38]

13 When the amount of left ventricular myocardium in a shock pathway is successful rhythm conversion is more likely. [Pg.38]


See other pages where Shock Polarity and Configuration is mentioned: [Pg.38]   