Quantitative DTA Heat-Flux DSC

Three kinds of sample holders are available for DTA and DSC (Fig. 4.8.6). Type 1 holders are commonly used for a DTA apparatus. In this type, the sample and reference holders are placed on the same metal block and heated by the same heater the temperature difference between the two holders is indicated by a thermocouple. Type 2 holders are generally used in a quantitative DTA (heat-flux DSC) apparatus. Both sample and reference holders are maintained at the same temperature by two individual internal heaters, which, in turn, are heated by the same main heater. The temperature difference between the sample and the reference material is measured by a thermocouple placed outside of the holders. Type 3 holders are customarily used for a power-compensated DSC apparatus. This apparatus has separate heaters for heating the sample and reference holders thus maintaining the sample and the reference... [Pg.204]