Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Peltier units

Figure 6 Schematic of an isothermal calorimeter unit showing the position of the reaction ampoule, Peltier units and heat sink. The instrument has different combination of application, including flow through or ampoule insertion (Courtesy Thermometric Ltd, Jafalla, Sweden)... Figure 6 Schematic of an isothermal calorimeter unit showing the position of the reaction ampoule, Peltier units and heat sink. The instrument has different combination of application, including flow through or ampoule insertion (Courtesy Thermometric Ltd, Jafalla, Sweden)...
The procedure used to study par tide/solidification front interactions is as follows. Matrix material and a small quantity of the test particles are placed in a groove machined In the upper face of a rectangular copper plate. Then the matrix material Is melted. By selectively heating or cooling each end of the copper cell by means of Peltier units, the solidification rate of the matrljc material Is controlled. The Interaction of particles with the advancing solidification front Is then observed through a microscope. ... [Pg.156]

Peltier, J.M. MacLean, D.B. Szarek, W.A. Determination of the Glycosidic Linkage in Peracetylated Disaccharides Comprised of D-Glucopyranose Units by Use of Desorption Electron-Ionization Mass Spectrometry. Rapid Commun. Mass Spectrom. 1991,5,446-449. [Pg.220]

Figure 5 Temperature-dependent turbidimetry profiles for solutions of poly(Lys-25) at pH 7.0 and 11.0. These measurements were performed on a Pharmacia Biotechnology Ultrospec 3000 UV/visible spectrophotometer equipped with a programmable Peltier cell and temperature control unit. Figure 5 Temperature-dependent turbidimetry profiles for solutions of poly(Lys-25) at pH 7.0 and 11.0. These measurements were performed on a Pharmacia Biotechnology Ultrospec 3000 UV/visible spectrophotometer equipped with a programmable Peltier cell and temperature control unit.
For mashed potato texture, profile analysis (TPA) and cone penetration tests are performed with a TA HDi Texture Analyser (Stable Micro Systems Ltd, Godaiming, UK). During the tests, the mashed potatoes are kept at 55° C by means of a temperature-controlled Peltier cabinet (XT/PC) coupled to a separate heat exchanger and PID control unit. For the cone penetration tests, a spreadability rig is used, consisting of a 45° conical perspex probe (P/45°C) that penetrates a conical sample holder containing 7 0.1 g of mashed potatoes (Alvarez et al., 2005 Canet et al., 2005b Fernandez et al, 2006). [Pg.176]

Zhou et al. [175] described the determination of severe acute respiratory syndrome (SARS) coronavirus by a microfluidic chip system. The unit included an LIF microfluidic chip analyzer, a glass microchip for both PCR and capillary electrophoresis, a chip thermal cycler based on dual Peltier thermoelectric elements, a reverse transcription-polymerase chain reaction (RT-PCR) SARS diagnostic kit, and a DNA electrophoretic sizing kit. According to the authors, the system allowed efficient DNA amplification of the SARS coronavirus followed by electrophoretic sizing and detection on the same chip. [Pg.225]

There are also two well-known thermoelectric effects resulting from the joining of dissimilar materials (forming a junction) the Seebech effect, on which thermocouples are based, and the Peltier effect, used for thermopiles. The Seebech effect results when the two junctions of the dissimilar materials are held at different temperatures. The Seebech coefficient, e, is defined as the open-circuit voltage generated per unit temperature differential of the two junctions ... [Pg.373]

The Peltier effect, on which thermopiles are based, is the heat released or taken up when a current flows through an isothermal junction. In a thermopile, the current flows in the opposite direction through the two junctions and, thus, transfers heat from one junction to the other. The Peltier coefficient, II, is defined as the heat absorbed per unit charge flow ... [Pg.373]

In 1834, Peltier observed that the passage of electric current/through a bimetallic circuit caused the absorption of heat at one junction and rejection of heat at the other junction. The heat flow per unit current at constant temperature was called the Peltier heat qPe, and defined by... [Pg.407]

Figure 7.5 shows a composed of a bimetallic couple metal wires a and b with one junction maintained at temperature T and the other maintained at T+ dT. An electromotive force E causes a current / to pass through the wires. A Peltier heat qpe(T + dT) per unit current will be absorbed at the warm junction and an amount of heat qpe(T) will be given off at the cool junction. To maintain a temperature gradient, Thomson heat (q l h i)(dT) must be supplied to the metal a, and an amount of heat (r/Th h)(c/7 j must be removed from b, since the current is in the opposite direction in metal wire b. In a closed work cycle, the electric energy is fully converted to heat. Therefore, the energy balance per unit current by the first law of thermodynamics is... [Pg.407]

Heating units are usually required to speed certain physico-chemical processes and normally consist of thermostated baths (either water baths or dry-type units such as the Peltier system) in which the robot places the tubes containing the reacting mixture and where these are kept for a preselected period. [Pg.260]

A DSC instrument contains two metal cells, sample and reference, enclosed within a thermal shield. The temperature of both is raised simultaneously with an external heater at a rate slow enough to ensure thermal and chemical equilibrium within the cells at all temperatures. Peltier thermoelectrie units associated with the sample cell maintain its temperature the same as the reference cell. The raw data is the heat flow from the Peltier device. The heat capacity for a protein as a function of temperature is typically linear below the unfolding transition (Fig. IB). As the temperature of the sample enters a range where an unfolding transition occurs, the sample takes on an anomalous (or excess) heat capacity since a portion of the heat transferred to the sample cell drives the transition. On completion of the transition... [Pg.393]

The Peltier coefficient has a simple physical meaning. It is the energy carried by the electrons per unit charge. The energy is measured relative to the Fermi energy. Each electron contributes to U in proportion to its relative contribution to the total conductivity. The weighting factor for electrons with dE at E is thus a(E)dE/a. [Pg.233]

Chemical reactions are temperature sensitive, and indeed, chemical rate constants and reactions mechanism are expected to vary considerably with temperature. Most investigations on the electrocatalysis of the ORR are usually performed at ambient conditions, which do not necessarily represent the behavior of the materials and the reaction at the conditions of practical interest. For example, in proton exchange membrane fuel cells, the temperature of operation is between 80 and 100 °C. Significant discrepancy in behavior may arise if reactions and materials are tested at ambient conditions and their behavior at high temperatures is merely deduced firom extrapolation. Schafer et al. introduced variable temperature SECM, with an operational range of 0-100 °C, by integrating a temperature control unit (Peltier element) into an SECM setup, as shown in the schematic of Fig. 23 [66]. At the heart of the temperature control unit is the Peltier element, which is housed in a stainless steel block. [Pg.130]

The design of the calorimeter is shown in Figure 9.19. The calorimeter consists of a burner and an electrical compensation heater in a double-walled heat pipe filled with Freon with a boiling point of 25 °C. The Freon transports the heat, which is developed by the burner and the compensation heater, to a cooling unit that consists of water-cooled Peltier elements. A thermometer located in the vapor phase is used to control the power of the compensation heater so as to keep the temperature of the vapor constant independent of the heat of combustion of the gas. The flow of gas burned in the calorimeter is determined by means of mass... [Pg.277]

Most instrument manufacturers will make temperature control units for use at around a few tens of degrees from ambient, with either thermoelectric (Peltier) temperature control, or more simply a cell holder which allows water, or other liquid, circulation from an external temperamre controlled bath. Cryostats are also available for precise very low temperature control, easily down to 77 K using liquid nitrogen cooling, less easily down to 4 K using liquid helium, and even to lower temperatures if required. A hot air blower, such as a hair dryer, is a convenient way to raise the temperature of a sample up to a few tens of degrees above ambient for the occasional experiment and a thermocouple or thermistor a convenient way to measure sample, or cell holder, temperature. [Pg.522]


See other pages where Peltier units is mentioned: [Pg.137]    [Pg.145]    [Pg.137]    [Pg.145]    [Pg.450]    [Pg.224]    [Pg.226]    [Pg.548]    [Pg.80]    [Pg.463]    [Pg.358]    [Pg.46]    [Pg.301]    [Pg.91]    [Pg.11]    [Pg.99]    [Pg.57]    [Pg.248]    [Pg.675]    [Pg.89]    [Pg.2183]    [Pg.2129]    [Pg.484]    [Pg.377]    [Pg.156]    [Pg.184]    [Pg.695]    [Pg.625]    [Pg.708]    [Pg.837]    [Pg.153]    [Pg.337]    [Pg.221]    [Pg.204]   
See also in sourсe #XX -- [ Pg.145 ]




SEARCH



Peltier

© 2024 chempedia.info