NMR Acquisition in Reaction Monitoring Stopped- and Continuous-flow

NMR Acquisition in Reaction Monitoring Stopped- and Continuous-flow [Pg.124]

The alternative method is continuous-flow , in which the reactants flow through the detection coil during data acquisition. Continuous-flow NMR techniques have been used for the direct observation of short-lived species in chemical reactions [4—6]. The main difference between stopped- and continuous-flow NMR is that in the latter the sample remains inside the detection coil only for a short time period, termed the residence time, x [7], which is determined by the volume of the detection cell and the flow rate. The residence time alters the effective relaxation times according to the relationship in Eq. (2.5.1)  [Pg.124]

It is also clear from Eq. (2.5.1) that the linewidth of the observed NMR resonance, limited by 1/T2, is significantly broadened at high flow rates. The NMR line not only broadens as the flow rate increases, but its intrinsic shape also changes. Whereas for stopped-flow the line shape is ideally a pure Lorentzian, as the flow rate increases the line shape is best described by a Voigt function, defined as the convolution of Gaussian and Lorentzian functions. Quantitative NMR measurements under flow conditions must take into account these line shape modifications. [Pg.125]




SEARCH



Continuous flow

Continuous monitoring

Continuous reaction monitoring

Continuous reactions

Continuous-flow reaction

Flow and reactions

Monitor reaction

NMR continued)

Reaction monitoring

Stop-flow

Stopped flow

Stopped-flow NMR

© 2024 chempedia.info