HPLC Troubleshooting Quick Reference

This section is designed to assist in troubleshooting system problems. It is not a systematic approach as described in Chapter 10, which always yields better results. Keeping this in mind, I have listed a series of commonly seen problems, possible causes, and suggested treatments. [Pg.221]

Problem li No power. (Display does not light up on module or system.) [Pg.221]

Treatment Check the plug at the socket and at the module. It may have worked loose or been unplugged accidentally. [Pg.221]

Cause b No fuse, blown fuse, or incorrect fuse. [Pg.221]

Treatment On a new system, make sure fuse(s) were installed. If variable voltage, make sure the selection card is correctly positioned. New systems are often shipped with fuses in a bag. If fuse is broken, replace it and contact service. [Pg.221]