El mass spectral libraries

Provided El spectra have been measured under some sort of standard conditions (70 eV, ion source at 150-250 °C, pressure in the order of 10 " Pa), they exhibit very good reproducibility. This is not only the case for repeated measurements on the same instrument, but also between mass spectrometers having different types of mass analyzers, and/or coming from different manufacturers. This property soon led to the collection of large El mass spectral libraries, either printed [76-78] or computerized. [79] The best established El mass spectral databases are the NIST/EPA/NIH Mass Spectral Database and the Wiley/NBS Mass Spectral Database, each of them giving access to about 120,000 evaluated spectra. [80-83]... [Pg.218]

Mass spectrometers can readily be interfaced to GC (Figure 4.21). As the compounds leave the chromatograph, they are introduced into the mass spectrometer operating in a vacuum. The great majority of GC-MS are benchtop instruments with linear quadrupoles and EL Compound identification is performed by comparison of the spectrum with a data base for precise identification and confirmation. Huge El mass spectral libraries such as NIST Library or Wiley Library contain... [Pg.216]


See also in sourсe #XX -- [ Pg.218 ]

See also in sourсe #XX -- [ Pg.242 ]