EB or BE geometry mass analysers

However, since the velocities of ions before acceleration are non-zero, their total kinetic energy must take account of their initial kinetic energy, Eq, which is small and unknown  [Pg.376]

Only the ions which have followed the trajectory with radius R within the flight tube imposed by the instrument will be detected. Hence, to obtain a spectrum within a given mass range, the intensity of the field B has to be progressively modified. All ions, whatever the miz value, will successively follow, one after the other, the unique trajectory leading to the detector. Such a scan requires around [Pg.378]


See other pages where EB or BE geometry mass analysers is mentioned: [Pg.374]    [Pg.375]    [Pg.377]   