Diffusion head sensor cell

Development of a Diffusion Head Sensor Cell. The use of air sampling pumps in portable electrochemical gas detection apparatus introduces potential problems into the instrument. First, the sensor cell response is dependent on gas flow rate. The sample flow rate, therefore, must be accurately controlled to obtain reproducible results, or the sample flow rate must be set high enough to insure a flow independent response. Secondly, failure of the pump itself could prevent a sample from reaching the sensor cell. Thirdly, the pumps are usually one of the largest users of current in a portable instrument and thereby limit usable battery life. [Pg.569]

The need for an air sampling pump can be eliminated by use of a diffusion tube having a set length to diameter (L/d) ratio in its geometry for introduction of a gas sample. Proper selection of the geometry and L/d ratio of the diffusion tube results in an electrochemical cell with a response which is independent of external gas flow rate. A schematic of a solid polymer electrolyte diffusion head sensor cell is shown in Figure 13. [Pg.569]

TEMPERATURE DEPENDENCE, DIFFUSION HEAD SENSOR CELL... [Pg.572]


See also in sourсe #XX -- [ Pg.569 , Pg.570 , Pg.571 ]