Adjustment of Pulse Frequency

For routine PFT NMR measurements, the pulse frequency is adjusted by changing the frequency of the transmitter until the CW spectrum of a reference sample with signals at both ends of the spectral width (e.g. acetone, Fig. 1.10) is reproduced exactly by Fourier transformation of the FID signal. [Pg.31]

If the pulse frequency offset is chosen in the center of Larmor frequency range zl, both halves A/2 of the spectrum are sampled simultaneously and folded into each other. Folding can be corrected by a modification of the Fourier transformation software and [Pg.31]


See other pages where Adjustment of Pulse Frequency is mentioned: [Pg.31]    [Pg.99]   